EMOCIA PH-W4710/4720

2262 Hits
Quantity in stock: Unlimited
Price:
0.00
Quantity:
-Standard Microwave Therapy Unit -Two Types: 1-Channel Type & 2-Channel Type -Various safety functions equipped. -Simple start in two steps. -2 treatment sounds, 4 ending sounds -The arms can be freely adjusted.

Simple Operation Panel

EMOCIA PH-W4720 Operation Panel

Output display

Timer display

Treatment time setting switch

Stop switch

Output volume


 

Simple Operation

Starting the treatment

1. Set the treatment time by treatment time seting switch.

2. Turn up the output volume to start the treatment.

 

Finishing the treatment

1. When the timer expires, the treatment is automatically finished.

*Please press the stop button, if you want to finish during the treatment.


 

Safety Function

1. Over output protection function

When the output exceeds 35% of the rating, the output automatically shuts off.

2. Output zero start

To prevent excessive irradiation, once the position is not returned to the zero position, the irradiation output cannot be adjusted.

3. Magnetron overheat prevention circuit

Even if the cooling fan of the magnetron breaks down, overheating is prevented by the thermostat.

4. Jam radio waves leakage prevention circuit

Even if the microwave from the irradiator is applied to the operation panel, it does not affect the circuit board.


 

EMOCIA Series

EMOCIA 4710(1-CH type)

EMOCIA 4720(2-CH type)

Specifications

 

EMOCIA 4710(1-CH)

EMOCIA 4720(2-CH)

Commodity code

PH-W4710AEK/CEK

PH-W4720BEK

Main power

AC100V or 220V 50/60Hz

AC100V or 220V 50/60Hz

Power consumption

500VA

1,000VA

Dimensions

450(W) x 690(H) x 300(D)mm

450(W) x 690(H) x 300(D)mm

Weight

30kg

43kg

Output

0-125W

0-125W

Frequency

2,450±50MHz

2,450±50MHz

Therapy modes

Continuous

Continuous

Treatment time

1-30 min

1-30 min

Safety function

Zero start, Over output protection function, Jam radio waves leakage prevention circuit, Magnetron overheat prevention circuit

Zero start, Over output protection function, Jam radio waves leakage prevention circuit, Magnetron overheat prevention circuit

Electrical class, type

Class I, Type B

Class I, Type B